FULLY AUTOMATIC
MICRO-PROCESSOR
AUTOCLAVE / STERILIZER
MODEL ALBERT - SERIES
FEATURES
Class B Type. Meets EN-13060
requirement.
Fully Automatic Microprocessor
Controlled.
Can be equip with an OPTIONAL interface Recorder
(RS232)
Temperature Control
Digital Display
Patented Design Dual Safety Door Lock (Pressure &
Vacuum Auto-Lock Device)
Dual Cut Off Protection Switch (Over
Heat &
Over
Pressure)
Unlock Door Alarm
Emergency Stop Exhaust Switch.
Dual Low Water Indication with Buzzer
Alarm. (Water
Reservoir & Chamber)
Prion Sterilization Mode.
Fast Sterilzation
Bowie Dick Test Mode
Re-Dry Mode (for Enhance Dry
Performance)
Date / Time Setting
Non-Operational Mode. (Check both of
Low Water &
Unlock Door)
SAFETY DEVICE
Dual Safety Door Lock (Patent Design)
Non-Operational Model of Device when door is not
properly closed and lock
Safety Valve
Pressure Protection Switch
Water Filling Detection Device
Door Unlock Reminder |
 
|
Low Water Warning Device triggers Buzzer
when water
in chamber is low
Emergency Exhaust Switch
Overheat / Over Pressure Indicator Light
STANDARD ACCESSORIES
Tray Set (Stainless Steel Rack + Aluminum Tray +
Stainless Steel Pull Holder) = 1 complete set
Stainless Steel Heating Element Cover
OPTIONAL ACCESSORY
RS-232 Printer |
TECHNICAL SPECIFICATIONS
| MODEL |
ALBERT-16D |
ALBERT-25D |
| CHAMBER SIZE |
(D)261 x (L)300mm. |
(D)261 x (L)470mm. |
| CAPACITY |
16 LITERS |
25 LITERS |
| OVERALL SIZE (W) x (D) x (H) |
525 x 691 x 425mm. |
525 x 861 x 425mm. |
| STERILIZATION MODE |
MODE |
TEMPERATURE |
PRESSURE |
STERILIZATION
TIME |
WRAPPED |
121°C |
1.05KG/CM2 |
30 MINUTES |
WRAPPED |
134°C |
2.07 KG/CM2 |
15 MINUTES |
UNWRAPPED |
121°C |
1.05 KG/CM2 |
15 MINUTES |
UNWRAPPED |
134°C |
2.07 KG/CM2 |
4 MINUTES |
BOWIE DICK
TEST |
134°C |
2.07 KG/CM2 |
4 MINUTES |
PRION |
135°C |
2.16 KG/CM2 |
18 MINUTES |
| DESIGN TEMPERATURE |
149°C |
| CONTROL UNIT |
MICROPROCESSOR TEMPERATURE CONTROLLER |
| AUTO WATER FILLING |
YES |
| AUTO EXHAUST FUNCTION |
YES |
| FAST STERILIZATION MODE |
YES |
| PRE-VACUUM |
YES |
| DRY FUNCTION (POST VACUUM) |
YES |
| DRY TIME (VACUUM) |
20 MINUTES (DEFAULT) |
| RE-DRY TIME |
10 MINUTES (DEFAULT) |
| VACUUM RELEASE |
YES |
| AIR FILTER |
YES |
| RECORDER |
RS232 INTERFACE RECORDER (OPTIONAL) |
| VACUUM PUMP |
DIAPHRAGM |
|